Introduction

The VIM-2 is a stationary (grid operation) and mobile, battery-operated,
highly accurate and long-term stable vacuum pressure gauge based on
Spinning Rotor Gauge (SRG) technology.

Spinning Rotor Gauges are characterized, among other things, by the option to separate the sensor and measuring head / electronics and thus enable checking static vacuums in a very cost-effective manner. Due to its simplicity, the all-metal sensor can be permanently welded to the vacuum system and delivers reliable measured values even after years of use. Only the ball in the tube, and no electronic components, remain on the chamber to be measured. The system is uniquely resistant to mechanical stress and non-chlorinated corrosive media. Any interaction with the vacuum chamber takes place via electromagnetic forces that lift and accelerate the ball and measure its braking behavior. Corresponding mechanical protection that needs to be attached from the outside is included.

Advantages – Autonomous, high-accuracy measurement, long-term stability

Spinning Rotor Gauges were previously considered to be complex to operate and were therefore almost exclusively used in calibration laboratories. A main application was, for example, the calibration of high vacuum measuring devices, ionization gauges.

The VIM-2, as a battery-operated device, is user-friendly, works largely independently and can be run remotely (App via Bluetooth). It is for instance particularly suitable for measurements on cryogenic, vacuum-insulated liquid gas tanks. It´s a safe option as it does not require any electrical feedthroughs, making it an all-metal sensor that is extremely robust and resistant to aging. Compared e.g., to the Pirani-based measuring method, the measuring range is extended by approx. 3 decades into the high vacuum region, and thus provides greater and far more valuable information. It is a direct, mathematical / physical method, that only uses the effect of decelerating a rotating ball as it collides with the gas molecules as a measure of pressure.

Most other methods use ionization use ionization by means of hot filaments or  high  voltage  to  indirectly  measure  the  pressure, they  are

mechanically less stable and sensitive to contamination, even impact the vacuum pressure measurement themselves (by degassing/ionization and sputtering). Heat transfer methods (Pirani) have a limited measurement range below 10-3 mbar (w/o zeroing, which is often impossible). In general, all other UHV measurement methods have electrical feedthroughs and fragile electronic components in the vacuum part.

SRGs on the other hand have a unique high level of accuracy, long-term stability, no feedthroughs, measure with a simple sphere at ambient temperature, in a simple tube, all interaction with the vacuum is through electromagnetic forces.

The power supply-independent VIM-2 gives total mobility for large and remote measurement locations, it runs for hours before recharge.

Despite of SRGs of the past, a reliable measurement is provided due to the strong magnetic fields and high rotation frequency of the sensor. VIM-2 generates a magnetic field that is twice as strong (90 mT) as previously available SRGs. The standard rotation frequency of the sphere at 615 Hz provides a much more stable suspension in the magnetic field. Both effects result in an improved tolerance to imperfect alignment of the sensor tube on e.g., the cryo vacuum tank and more reliably capturing the sensor ball.

The measuring heads are not sensor-specific and therefore interchangeable with one another, and previously installed sensors, so each head can measure a large number of sensors without a significant loss of accuracy.

Measurement principle

The Spinning Rotor Gauge is based on the effect that gas molecules slow down a moving body as they are adsorbed on its surface and then re-emitted. The statistically distributed molecules striking from all directions are accelerated accordingly and break the body (momentum transfer, area-related particle flux density).

In the low-pressure regime (< 1 x 10-7 mbar) the desorption of the accelerated gas molecules does not contribute to the rotation frequency. At pressures above 1 x 10-4 mbar and mean free paths in the cm range, the gas viscosity is automatically compensated by means of an integrated temperature measurement.

In the case of the SRG, a simple precision ball is used as the sensor and the deceleration is determined by inductive measurement in the sensor head. Factors such as the molecular mass and temperature changes are only included in the measurement in an easily determinable root ratio.

Additionally, within small limits, variable factors determine the braking of the rotating ball:

  1. The “residual drag” results from the material properties of the ball. It is particularly relevant for < 10-5 mbar and can be roughly pre-determined and stored as required.
  2. The “molecular drag” is the actual deceleration by the surrounding medium and proportional to the pressure. The molecular drag contains the calibration factor, which deviates from a ideally smooth sphere (factor 1) by max. ± 6 % without any calibration, ± 1% can be achieved.

It should be noted that even without setting the calibration factor or the zero point, and even after years, the system still has a unique intrinsic accuracy in high vacuum, which cannot be achieved by any other vacuum measurement method and is more than sufficient for many applications.

 

Measurement range

Due to their principle, Spinning Rotor Gauges can be used in pressure ranges from atmosphere to UHV. Due to geometric and electromechanical conditions, however, they are optimized for the pressure range, in which they achieve uniquely high levels of accuracy and long-term stability:

1% accuracy of the measured value between 0.1 and 1 x 10-6 mbar can be achieved if needed, <1% drift per year is typically. The measuring range in general is 10 mbar to 5 x 10-7 mbar, higher pressure values on request.

VIM-2 Specifications
Measurement PrincipleSpinning Rotor Gauge
Measurement range7 Torr to 4 x 10-7 Torr
10 mbar to 5 x 10-7 mbar
Achievable accuracy7 Torr to 0.07 Torr
10 mbar to 0.1 mbar approx. 10 % of reading
0,07 Torr to 7 x 10-7 Torr
0,01 mbar to 1 x 10-6 mbar 1 % of reading
Long term stability< 1 % drift per year
Wetted Material tube/ball1.4404, 1.4034
AlignmentHorizontal
Analog InterfacesUSB Type C, Bluetooth, High Density 15-polig
Digital InterfacesUSB, Bluetooth, RS 232, RS 485
Remote Control /GUIWin7/10
Operating TemperatureElectronics 10 °C to 40 °C (50 F to 104 F)
Measurement head 10 °C to 50 °C (50 F to 122 F)
Sensor tube bakeable until
150 °C (302F) elastomer sealed
450 °C (842F) metal sealed
Power Requirements18 to 28 VDC
Analog Output0 - 10 VDC configurable
Flange TypeKF 25, 8 VCR, welding socket
WeightDevice 1650 g (58 oz), power adapter 510 g (18 oz), head 750 g (26.5 oz)
DimensionsL x B x H = 325 x 275 x 100 mm (13 x 11 x 4 in.)
Power supplyGrid or battery operation
Power ConsumptionMaximum 1.5 A at 24 VDC, typically < 1 A
Run time in battery mode> 2 h
Battery charging timeIn 4.5 h to more than 95 % of current capacity